File size: 35,862 Bytes
a652f65 |
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127 128 129 130 131 132 133 134 135 136 137 138 139 140 141 142 143 144 145 146 147 148 149 150 151 152 153 154 155 156 157 158 159 160 161 162 163 164 165 166 167 168 169 170 171 172 173 174 175 176 177 178 179 180 181 182 183 184 185 186 187 188 189 190 191 192 193 194 195 196 197 198 199 200 201 202 203 204 205 206 207 208 209 210 211 212 213 214 215 216 217 218 219 220 221 222 223 224 225 226 227 228 229 230 231 232 233 234 235 236 237 238 239 240 241 242 243 244 245 246 247 248 249 250 251 252 253 254 255 256 257 258 259 260 261 262 263 264 265 266 267 268 269 270 271 272 273 274 275 276 277 278 279 280 281 282 283 284 285 286 287 288 289 290 291 292 293 294 295 296 297 298 299 300 301 302 303 304 305 306 307 308 309 310 311 312 313 314 315 316 317 318 319 320 321 322 323 324 325 326 327 328 329 330 331 332 333 334 335 336 337 338 339 340 341 342 343 344 345 346 347 348 349 350 351 352 353 354 355 356 357 358 359 360 361 362 363 364 365 366 367 368 369 370 371 372 373 374 375 376 377 378 379 380 381 382 383 384 385 386 387 388 389 390 391 392 393 394 395 396 397 398 399 400 401 402 403 404 405 406 407 408 409 410 411 412 413 414 415 416 417 418 419 420 421 422 423 424 425 426 427 428 429 430 431 432 433 434 435 436 437 438 439 440 441 442 443 444 445 446 447 448 449 450 451 452 453 454 |
---
license: apache-2.0
language:
- en
pipeline_tag: summarization
widget:
- text: 'the need for magnetic resonance imaging ( mri ) in patients with an implanted
pacemaker or implantable cardioverter - defibrillator ( icd ) is a growing clinical
issue . it is estimated that as many as 75% of active cardiac device recipients
will become indicated for mri . currently , the vast majority of such devices
are contraindicated for use with an mri . in european heart rhythm association
survey , published recently for non - mri - certified icds ( 0.51.5 t field strength
) , the totally subcutaneous icd ( s - icd ) system , an implantable defibrillator
with no leads that touch the heart , has recently been demonstrated to be a safe
and effective defibrillator option for patients at risk for sudden cardiac death
. it provides shock therapy and post - shock pacing therapy , but no long - term
bradycardia pacing . although it has been shown as an alternative to the standard
transvenous icd , its compatibility with mri remains unclear . various types of
clinical mri systems currently use a superconductive magnet that creates a static
magnetic field strength , typically 1.5 or 3 t. the use of mri with most pacemakers
and icds is considered a contraindication due to potential hazards , including
heating of the electrode that resides in or on the heart , damage to myocardium
, elevation of pacing thresholds , unintended induction of ventricular tachycardia
( vt ) or ventricular fibrillation ( vf ) , pacing inhibition , permanent device
malfunction , and distortion of the mri scan . recently , mr - conditional. mr
- conditional indicates a lack of known hazards in a specified mri environment
with specified conditions of use . due to the variety of mri scanners and scanning
protocols , it is not practical to test even a single device under all conditions
. hence , mr - conditional labelling dictates that the device is safe for use
under certain scanning conditions , as well as how the cardiac device should be
programmed before an exposure to the magnetic field in a mri scanner . the literature
, although limited , provides some guidance for imaging patients with implanted
pacemakers or icds that do not have mr - conditional labelling . this single -
centre prospective non - controlled study describes the first use of mri in patients
with an implanted s - icd . patients with implanted s - icd systems ( boston scientific
sqrx model 1010 and q - trak model 3010 ) were enrolled for mri testing over a
period of 18 months . the s - icd system implanted in this patient cohort was
composed of a can implanted in a left mid - lateral pocket and a para - sternal
subcutaneous electrode . the s - icd is currently not certified for use with an
mri ; therefore , the ethics committee of homolka hospital , prague , czech republic
approved our clinical study . patients with newly implanted s - icd systems (
< 6 weeks ) were excluded , and none of the patients had any intravascular leads
. the patients were randomized for either a cardiac , brain , cervical , or lumbar
spinal scan . one of the subjects underwent an additional knee examination , due
to reported chronic pain . a total of 15 patients were enrolled into this study
( 12 males and three females , aged 2283 years , mean 53 years . subjects in our
cohort ( table 1 ) underwent a total of 22 mri scans between 6 june 2012 and 24
december 2013 . in total , five brain scans , three cardiac scans , 12 lumbar
scans , one knee , and one cervical spine scan were conducted ( table 2 ) . however
, in one patient a minor disc protrusion was found , in other mri revealed stenosis
of intervertebral foramen which was causing radicular pain of the nerve root l4
and based on this examination the patient was referred to ct - navigated periradicular
therapy . table 1summary of patient anatomical data and scan locations , along
with noted clinical eventsidagesexbmidgef , % indication for s - icdheating0164f20.5hcmp
/ vfs85secondary preventionnone0283m30.0post - mi / smvts post - catheter ablation/35secondary
prevention ( post - transvenous icd extraction)none0331m25.3arvc / d / smvts68secondary
preventionin - tolerable re - scanned0458m23.6post - mi / post - cabg30primary
preventionnone0577m25.5post - mi30primary preventionnone0663m27.0post - mi30primary
preventionnone0768m23.7post - mi / vfs / vts60secondary prevention post - transvenous
icd extraction / svc occlusiontolerable0822m29.4brugada sy / vfs68secondary preventionin
- tolerable re - scanned0959m27.1dcmp / vfs / post - mitral valve surgery/60secondary
prev./post - transvenous icd extractionnone1041f24.6arvc / d70primary preventionnone1123f21.5lqts
/ vf60secondary preventionnone1266m36.9post - mi / vf / post - cabg50secondary
prevention / post - repeat transvenous icd extractiontolerable1348m22.9dcmp(non
- compaction)/vfs35secondary preventionnone1470m29systolic dysfunction of lv35primary
preventionnone1526m33brugada sy65primary preventionnonehcmp , hypertrophic cardiomyopathy
; smvt , sustained monomorphic ventricular tachycardia ; mi , myocardial infarction
; arvc , arrhythmogenic right ventricular cardiomyopathy ; cabg , coronary artery
by - pass graft ; lqts , long qt syndrom . table 2parmeters of s - icd and patient
sensation during individual mri scansscan # idbody partheating sensationsshock
zone ( b.p.m.)condit . shock zone ( b.p.m.)bat % episode num.101brainnone2302101001202brainnone240220861303l
spinein - tolerable240220831403brainnone240220831504brainnone220190691605l spinenone220210541706l
spinenone240220681807l spinetolerable240220582908l spinein - tolerablenananana1008brainnonenananana1108l
spinenone2302108411209heartnone2402208911310l spinenone2301807911410heartnonenananana1511heartnone2301909711612l
spinetolerable2001709721712l spinenone2001709421813c spinenone23019010041913l
spinenone23019010042014l spinenone2301908612115kneenone25021010012215l spinenone2502101001s
- icd parameters acquired prior- and post - mri were without any change , therefore
only one value is presented.indices : na , not available ; l spine , lumbar spine
; c spine , cervical spine . summary of patient anatomical data and scan locations
, along with noted clinical events hcmp , hypertrophic cardiomyopathy ; smvt ,
sustained monomorphic ventricular tachycardia ; mi , myocardial infarction ; arvc
, arrhythmogenic right ventricular cardiomyopathy ; cabg , coronary artery by
- pass graft ; lqts , long qt syndrom . parmeters of s - icd and patient sensation
during individual mri scans s - icd parameters acquired prior- and post - mri
were without any change , therefore only one value is presented . indices : na
, not available ; l spine , lumbar spine ; c spine , cervical spine . studies
were performed using a siemens avanto 1.5 t mri scanner ( vb17 software , quantum
gradient coils ) . all scans were run in normal operating mode , which is limited
to 2 w / kg whole body averaged specific absorption rate ( sar ) . clinically
relevant mri sequences were used for evaluation ( see table 3 ) . table 3types
of pulse sequences typically used for imaging of respective anatomical areasscan
locationscan sequencesflairdwiflashfsehastesestirtruefispbrainxxxxheartxxxxcervical
spinexxkneexxxxlumbar spinexxflair , fluid attenuated inversion recovery ; dwi
, diffusion weighted imaging ; flash , fast low angle shot ; fse , fast spin echo
; haste , half acquisition single - shot turbo spin echo ; se , spin echo ; stir
, short tau inversion recovery ; truefisp , true fast imaging with steady - state
precession.fse sequence caused heating in subjects with a thermistor probe during
lumbar spine examination ( see the text for details ) . types of pulse sequences
typically used for imaging of respective anatomical areas flair , fluid attenuated
inversion recovery ; dwi , diffusion weighted imaging ; flash , fast low angle
shot ; fse , fast spin echo ; haste , half acquisition single - shot turbo spin
echo ; se , spin echo ; stir , short tau inversion recovery ; truefisp , true
fast imaging with steady - state precession . fse sequence caused heating in subjects
with a thermistor probe during lumbar spine examination ( see the text for details
) . patients were asked to report immediately any pain , torqueing movement ,
or heating sensation in the area of the pocket or the electrode by pressing an
emergency bulb . furthermore , all patients were questioned immediately following
the mri procedure to ascertain any discomfort in the vicinity of the can or electrode
. pulse oximetry and standard lead electrocardiogram ( ecg ) if discomfort occurred
, the patient was asked if the scan could be repeated at a later time using a
revised scan sequence or the subject was again randomized for another anatomical
area . since none of the components of the s - icd system are on or in the heart
, heating near or around however , heating near the electrode or can with the
s - icd system may still cause serious patient discomfort . therefore , along
with education of subjects , each patient was instrumented by taping an oesophageal
temperature probe ( beta - therm model g22k7mcd8 ) on the skin over the mid -
lateral implant site to record any temperature excursions that might be correlated
to patient symptoms of heating / discomfort near the pocket . to minimize the
risk of inappropriate therapy , the s - icd system was programmed to therapy each
s - icd system was evaluated prior to and immediately after the scan to verify
proper functioning , including interrogation , sensing , and battery voltage .
after the completion of the mri , long - term regular clinical follow - up and
checking of the device were performed . patients with implanted s - icd systems
( boston scientific sqrx model 1010 and q - trak model 3010 ) were enrolled for
mri testing over a period of 18 months . the s - icd system implanted in this
patient cohort was composed of a can implanted in a left mid - lateral pocket
and a para - sternal subcutaneous electrode . the s - icd is currently not certified
for use with an mri ; therefore , the ethics committee of homolka hospital , prague
, czech republic approved our clinical study . patients with newly implanted s
- icd systems ( < 6 weeks ) were excluded , and none of the patients had any intravascular
leads . the patients were randomized for either a cardiac , brain , cervical ,
or lumbar spinal scan . one of the subjects underwent an additional knee examination
, due to reported chronic pain . a total of 15 patients were enrolled into this
study ( 12 males and three females , aged 2283 years , mean 53 years . subjects
in our cohort ( table 1 ) underwent a total of 22 mri scans between 6 june 2012
and 24 december 2013 . in total , five brain scans , three cardiac scans , 12
lumbar scans , one knee , and one cervical spine scan were conducted ( table 2
) . however , in one patient a minor disc protrusion was found , in other mri
revealed stenosis of intervertebral foramen which was causing radicular pain of
the nerve root l4 and based on this examination the patient was referred to ct
- navigated periradicular therapy . table 1summary of patient anatomical data
and scan locations , along with noted clinical eventsidagesexbmidgef , % indication
for s - icdheating0164f20.5hcmp / vfs85secondary preventionnone0283m30.0post -
mi / smvts post - catheter ablation/35secondary prevention ( post - transvenous
icd extraction)none0331m25.3arvc / d / smvts68secondary preventionin - tolerable
re - scanned0458m23.6post - mi / post - cabg30primary preventionnone0577m25.5post
- mi30primary preventionnone0663m27.0post - mi30primary preventionnone0768m23.7post
- mi / vfs / vts60secondary prevention post - transvenous icd extraction / svc
occlusiontolerable0822m29.4brugada sy / vfs68secondary preventionin - tolerable
re - scanned0959m27.1dcmp / vfs / post - mitral valve surgery/60secondary prev./post
- transvenous icd extractionnone1041f24.6arvc / d70primary preventionnone1123f21.5lqts
/ vf60secondary preventionnone1266m36.9post - mi / vf / post - cabg50secondary
prevention / post - repeat transvenous icd extractiontolerable1348m22.9dcmp(non
- compaction)/vfs35secondary preventionnone1470m29systolic dysfunction of lv35primary
preventionnone1526m33brugada sy65primary preventionnonehcmp , hypertrophic cardiomyopathy
; smvt , sustained monomorphic ventricular tachycardia ; mi , myocardial infarction
; arvc , arrhythmogenic right ventricular cardiomyopathy ; cabg , coronary artery
by - pass graft ; lqts , long qt syndrom . table 2parmeters of s - icd and patient
sensation during individual mri scansscan # idbody partheating sensationsshock
zone ( b.p.m.)condit . shock zone ( b.p.m.)bat % episode num.101brainnone2302101001202brainnone240220861303l
spinein - tolerable240220831403brainnone240220831504brainnone220190691605l spinenone220210541706l
spinenone240220681807l spinetolerable240220582908l spinein - tolerablenananana1008brainnonenananana1108l
spinenone2302108411209heartnone2402208911310l spinenone2301807911410heartnonenananana1511heartnone2301909711612l
spinetolerable2001709721712l spinenone2001709421813c spinenone23019010041913l
spinenone23019010042014l spinenone2301908612115kneenone25021010012215l spinenone2502101001s
- icd parameters acquired prior- and post - mri were without any change , therefore
only one value is presented.indices : na , not available ; l spine , lumbar spine
; c spine , cervical spine . summary of patient anatomical data and scan locations
, along with noted clinical events hcmp , hypertrophic cardiomyopathy ; smvt ,
sustained monomorphic ventricular tachycardia ; mi , myocardial infarction ; arvc
, arrhythmogenic right ventricular cardiomyopathy ; cabg , coronary artery by
- pass graft ; lqts , long qt syndrom . parmeters of s - icd and patient sensation
during individual mri scans s - icd parameters acquired prior- and post - mri
were without any change , therefore only one value is presented . indices : na
, not available ; l spine , lumbar spine ; c spine , cervical spine . studies
were performed using a siemens avanto 1.5 t mri scanner ( vb17 software , quantum
gradient coils ) . all scans were run in normal operating mode , which is limited
to 2 w / kg whole body averaged specific absorption rate ( sar ) . clinically
relevant mri sequences were used for evaluation ( see table 3 ) . table 3types
of pulse sequences typically used for imaging of respective anatomical areasscan
locationscan sequencesflairdwiflashfsehastesestirtruefispbrainxxxxheartxxxxcervical
spinexxkneexxxxlumbar spinexxflair , fluid attenuated inversion recovery ; dwi
, diffusion weighted imaging ; flash , fast low angle shot ; fse , fast spin echo
; haste , half acquisition single - shot turbo spin echo ; se , spin echo ; stir
, short tau inversion recovery ; truefisp , true fast imaging with steady - state
precession.fse sequence caused heating in subjects with a thermistor probe during
lumbar spine examination ( see the text for details ) . types of pulse sequences
typically used for imaging of respective anatomical areas flair , fluid attenuated
inversion recovery ; dwi , diffusion weighted imaging ; flash , fast low angle
shot ; fse , fast spin echo ; haste , half acquisition single - shot turbo spin
echo ; se , spin echo ; stir , short tau inversion recovery ; truefisp , true
fast imaging with steady - state precession . fse sequence caused heating in subjects
with a thermistor probe during lumbar spine examination ( see the text for details
) . patients were asked to report immediately any pain , torqueing movement ,
or heating sensation in the area of the pocket or the electrode by pressing an
emergency bulb . furthermore , all patients were questioned immediately following
the mri procedure to ascertain any discomfort in the vicinity of the can or electrode
. pulse oximetry and standard lead electrocardiogram ( ecg ) if discomfort occurred
, the patient was asked if the scan could be repeated at a later time using a
revised scan sequence or the subject was again randomized for another anatomical
area . since none of the components of the s - icd system are on or in the heart
, heating near or around the electrode can not harm the myocardium . however ,
heating near the electrode or can with the s - icd system may still cause serious
patient discomfort . therefore , along with education of subjects , each patient
was instrumented by taping an oesophageal temperature probe ( beta - therm model
g22k7mcd8 ) on the skin over the mid - lateral implant site to record any temperature
excursions that might be correlated to patient symptoms of heating / discomfort
near the pocket . to minimize the risk of inappropriate therapy , the s - icd
system was programmed to therapy each s - icd system was evaluated prior to and
immediately after the scan to verify proper functioning , including interrogation
, sensing , and battery voltage . after the completion of the mri , the s - icd
system was reprogrammed to original settings . long - term regular clinical follow
- up and checking of the device were performed . no anomalies were noted via pulse
oximetry or ecg during the scans for any of the patients . eleven of 15 patients
reported no sensation or pain from heating of the can , two of 15 patients reported
feeling some heating , and two patients reported intolerable heating ( see table
2 ) . in patients with intolerable heating , the scan was halted within seconds
and changed to a scan of the brain , which proceeded without incident . patient
reports of heating in the vicinity of the can occurred only during lumbar scans
with a thermistor probe ; no such reports occurred during scans of the brain ,
cardiac area , cervical spine , or without the probe . in two cases where heating
in the vicinity of the can was reported by the patient , the scan sequence was
altered to reduce the intensity of radiofrequency ( rf ) field exposure by reducing
the turbo factor ( e.g. from 21 to 7 ) , increasing the repetition time ( e.g.
to > 4000 ms ) , and reducing the flip angle ( e.g. from 170 to 120 ) . the target
values were chosen arbitrarily to maintain image contrast ( flip angle ) and keep
scan time at reasonable limits ( turbo factor and repetition time ) . less heating
was noted by patients after these modifications to the scan parameters were made
. 03 ) was observed to have a skin lesion , appearing to be a circular rash or
ulcer on the surface of the skin over the can , approximately 35 mm in diameter
. the cause of this skin anomaly is not known ; it was later noted to have fully
healed at a follow - up 10 days after the scan . to ascertain the effect of heating
due to the instrumented thermistor catheter , the two patients who experienced
the heating ( examinations 9 and 16 , see table 2 ) were rescanned several weeks
later without the thermistor catheter in place ( examinations 11 and 17 ) . first
, modified sequence ( with even lower amount of energy deposited in the tissue
) was used , which caused no heating . as no sensation was reported by the subjects
, they were asked to report even a minimal discomfort , and the lumbar scans were
performed using the same settings that resulted in heating with the thermistor
catheter in place in the first imaging session . the results of the rescans revealed
that no heating was felt by the patients when the thermistor catheter was absent
. there were no noted changes to battery voltage , ability to detect the qrs signal
or stored diagnostic data . pacing thresholds can not be assessed by the s - icd
system , so this was not evaluated . none of the patients reported any pulling
or twisting of the can or pain from heating of the s - icd electrode . for scans
of the brain , lumbar spine , knee , and cervical spine , no effect from image
artefact was noted in the anatomical area of interest . however , for scans of
the cardiac area , image artefact was noted to interfere with the ability to see
parts of the left ventricle , though the right ventricle of the heart was unaffected
and could be imaged usefully . this was due to the can and not the electrode (
see figure 1 ) , modifications to the protocol for the lumbar spine resulted in
a lower signal - to - noise ratio ; however , the images remain in diagnostic
quality ( see figure 2 ) . figure 1kinetic study in four - chamber view : the
systolic ( a and c ) and diastolic ( b and d ) images of cine sequences , four
- chamber view . the steady - state free precession ( ssfp ) sequence ( a and
b ) shows more artefacts . in ssfp kinetic study , an inflow of dark blood from
the left pulmonary veins was seen . it could be caused by s - icd but also by
metallic ring in mitral annulus . the spoiled gradient echo ( gre ) sequence (
c and d ) is better , but an artefact at the lateral wall is obvious . figure
2lumbar spine imaging with icd : low sar t2 fse sequence ( upper image ) compared
with normal t2 fse in the same subject ( lower image , for the scanning parameters
see the discussion section ) . kinetic study in four - chamber view : the systolic
( a and c ) and diastolic ( b and d ) images of cine sequences , four - chamber
view . the steady - state free precession ( ssfp ) sequence ( a and b ) shows
more artefacts . in ssfp kinetic study , an inflow of dark blood from the left
pulmonary veins was seen . it could be caused by s - icd but also by metallic
ring in mitral annulus . the spoiled gradient echo ( gre ) sequence ( c and d
) is better , but an artefact at the lateral wall is obvious . lumbar spine imaging
with icd : low sar t2 fse sequence ( upper image ) compared with normal t2 fse
in the same subject ( lower image , for the scanning parameters see the discussion
section ) . there were no noted changes to battery voltage , ability to detect
the qrs signal or stored diagnostic data . pacing thresholds can not be assessed
by the s - icd system , so this was not evaluated . none of the patients reported
any pulling or twisting of the can or pain from heating of the s - icd electrode
. for scans of the brain , lumbar spine , knee , and cervical spine , no effect
from image artefact was noted in the anatomical area of interest . however , for
scans of the cardiac area , image artefact was noted to interfere with the ability
to see parts of the left ventricle , though the right ventricle of the heart was
unaffected and could be imaged usefully . this was due to the can and not the
electrode ( see figure 1 ) , modifications to the protocol for the lumbar spine
resulted in a lower signal - to - noise ratio ; however , the images remain in
diagnostic quality ( see figure 2 ) . figure 1kinetic study in four - chamber
view : the systolic ( a and c ) and diastolic ( b and d ) images of cine sequences
, four - chamber view . the steady - state free precession ( ssfp ) sequence (
a and b ) shows more artefacts . in ssfp kinetic study , an inflow of dark blood
from the left pulmonary veins was seen . it could be caused by s - icd but also
by metallic ring in mitral annulus . the spoiled gradient echo ( gre ) sequence
( c and d ) is better , but an artefact at the lateral wall is obvious . figure
2lumbar spine imaging with icd : low sar t2 fse sequence ( upper image ) compared
with normal t2 fse in the same subject ( lower image , for the scanning parameters
see the discussion section ) . kinetic study in four - chamber view : the systolic
( a and c ) and diastolic ( b and d ) images of cine sequences , four - chamber
view . the steady - state free precession ( ssfp ) sequence ( a and b ) shows
more artefacts . in ssfp kinetic study , an inflow of dark blood from the left
pulmonary veins was seen . it could be caused by s - icd but also by metallic
ring in mitral annulus . the spoiled gradient echo ( gre ) sequence ( c and d
) is better , but an artefact at the lateral wall is obvious . lumbar spine imaging
with icd : low sar t2 fse sequence ( upper image ) compared with normal t2 fse
in the same subject ( lower image , there are several reports in the current literature
about mr - conditional pacemakers from several companies , but very limited reports
about mr - conditional icds . biotronik announced in late 2011 release of their
first mr - compatible icd device and defibrillator leads pro mri , but in the
conditions of use excluded scanning of the torso and focused more on the extremities
examination . in european heart rhythm association survey , 60% of centres did
not implant any mri - certified icds , 34.3% implanted < 10 icd devices , and
only 5.6% implanted 10 and more icds ; one - fifth of responders stated that mri
- certified icds should be implanted in all patients but lack of reimbursement
was indicated as a possible obstacle to implant more mri - certified pacemakers
/ icds by 47.1% of responding centres . none of the components of the s - icd
system are on or in the heart . the s - icd depends less upon being in direct
contact with the myocardium to function and instead uses far - field sensing and
stimulation to provide the shock and post - shock pacing therapy . as a consequence
, unlike transvenous systems heating near or around the electrode can not harm
the myocardium , which could present with possible safety consequences such as
an elevation in pacing thresholds or scarring of the myocardium , but it may still
cause serious patient discomfort . because the s - icd is larger than modern transvenous
icd ''s , there may be more potential for the can to experience heating due to
the magnetic gradient or rf field . we report results from what we believe is
the first experience of mri scanning in patients with an implanted s - icd and
in various anatomical areas . overall , mri was performed safely in all patients
, which is in contrast to the current literature with mri imaging in patients
with electrical - active devices which are not mri - conditional . in our study
, the primary clinically significant event attributable to the mri scan was the
occurrence of heating in the area of the pocket in the four patients that underwent
lumbar scans . it was not known if this was due to the s - icd can itself or an
artefact of the thermistor catheter used to measure skin temperature over the
pocket . this required a revision of our protocol , which was to re - scan two
of the patients who complained of heating . re - scanning of these patients without
the thermistor probe resulted in no complaints of heating , so it is assumed that
the thermistor catheter itself heated during the lumbar scans and caused the discomfort
. as further evidence , all the heating complaints occurred during rf - intensive
scan sequences ( namely fast spin echo ) with the temperature probe located axially
near the centre of the bore , where rf fields are the highest . the thermistor
catheter is constructed of insulated conductive cables connected to electrodes
and should couple to the rf fields efficiently , causing heating at the electrodes
and pain or damage on the surface of the skin where the probe was placed over
the s - icd can . if the heating was due to the s - icd can itself , it would
more likely occur during gradient - intensive scan sequences ( which can generate
eddy currents on can surfaces and internal components ) and at locations in the
bore where there are high gradient fields , such as near the bore edges . however
, when the patient was scanned with gradient - intensive scan sequences ( e.g.
flair dwi ) and with the s - icd system in high gradient field locations in the
bore ( e.g. such as during a brain scan ) , patients did not detect any heating
or discomfort . in addition , the subcutaneous lead , which was not instrumented
with a thermistor catheter , never resulted in any heating sensation noted by
the patient , even when exactly the same sequence that resulted in heating in
the first session was used . the use of mri - compatible temperature monitors
such as fibre optic temperature probes would have provided better confirmation
of possible skin temperature elevation and would not have been affected by the
rf fields . for cardiac imaging , the main problem to solve is metallic artefact
, especially on the gradient - echo sequences . like in research performed by
nazarian et al . , several scan protocols were used to see if any yielded different
effects or reduced the qualitative extent of artefact . gradient mode was changed
from normal to whisper , resulting in slower ramping of the field and therefore
diminishing the changes of the magnetic field in time . artefacts when present
were limited to blurring of the left ventricle during cardiac scans and most yielded
clinically useful information . standard interrogation of the s - icd revealed
no adverse effects upon the functioning of the system . while no adverse effects
upon the post - scan s - icd device function were noted , not all possible scanning
protocols were tested . it should be noted that , four of the s - icd ''s were
exposed to repeat mri scans without adverse effects to device function . in addition
, because the s - icd does not provide long - term bradycardia pacing , it is
assumed that pacemaker - dependent patients would not be implanted with this system
. the inhibition of the pacemaker function during the scanning sequence and possible
pacing threshold changes are a unique concern in patients implanted with transvenous
icds . this study included only 15 patients and 22 scans done on the same 1.5
t mri scanner . thus , even these preliminary results should only be applied to
1.5 t mri scanners ( similarly as reported in the present literature for other
implantable devices ) . device functionality was tested immediately after the
scan but not for long - term effects . in addition , not all device functions
were tested although the s - icd system does have a beeper / interrogation warning
if battery levels or memory irregularities occur . however , patients were scheduled
for regular check - up and no defect of the device was observed in following 725
months ( mean observation time 18 months ) . delayed enhancement mri for determining
cardiac scarring was also not tested . also , there are other anatomical areas
that were not evaluated , such as shoulder and knees . this study included only
15 patients and 22 scans done on the same 1.5 t mri scanner . thus , even these
preliminary results should only be applied to 1.5 t mri scanners ( similarly as
reported in the present literature for other implantable devices ) . device functionality
was tested immediately after the scan but not for long - term effects . in addition
, not all device functions were tested although the s - icd system does have a
beeper / interrogation warning if battery levels or memory irregularities occur
. however , patients were scheduled for regular check - up and no defect of the
device was observed in following 725 months ( mean observation time 18 months
) . delayed enhancement mri for determining cardiac scarring was also not tested
. also , there are other anatomical areas that were not evaluated , such as shoulder
and knees . while more data are required to support a claim of mri - conditional
, this study is the study to demonstrate the feasibility of exposing s - icd patients
to mri using the scanning and monitor protocol described , with some precautionary
measures including : ( i ) programming the device therapy off ; ( ii ) limiting
the sar to 2.0 w / kg ; ( iii ) continuous monitoring of the patients pulse oximetry
and ecg by qualified personnel and especially for any feelings of heating ; (
iv ) evaluate device function post scan ; ( v ) availability of full resuscitation
facilities at the mri site . given the variables of different mri scanners , the
decision to perform mri on patients with an implanted s - icd system should be
balanced against the potential risks . in our study , the only heating was very
likely introduced by not fully mri - compatible thermometer probe ; subjects rescanned
without the probe did not report any abnormalities during the scan of any body
area listed ( brain , cervical and lumbar spine , heart , and knee ) . this study
was supported by iga mz r nt12094/2011 , research project charles university in
prague , prvouk p34 and unce 204010/2012 . funding to pay the open access publication
charges for this article was provided by iga mz r nt12094/2011 .'
example_title: Summarization Example 1
tags:
- medical
- llama-cpp
- gguf-my-repo
base_model: Falconsai/medical_summarization
---
# fernandoruiz/medical_summarization-Q4_0-GGUF
This model was converted to GGUF format from [`Falconsai/medical_summarization`](https://huggingface.co/Falconsai/medical_summarization) using llama.cpp via the ggml.ai's [GGUF-my-repo](https://huggingface.co/spaces/ggml-org/gguf-my-repo) space.
Refer to the [original model card](https://huggingface.co/Falconsai/medical_summarization) for more details on the model.
## Use with llama.cpp
Install llama.cpp through brew (works on Mac and Linux)
```bash
brew install llama.cpp
```
Invoke the llama.cpp server or the CLI.
### CLI:
```bash
llama-cli --hf-repo fernandoruiz/medical_summarization-Q4_0-GGUF --hf-file medical_summarization-q4_0.gguf -p "The meaning to life and the universe is"
```
### Server:
```bash
llama-server --hf-repo fernandoruiz/medical_summarization-Q4_0-GGUF --hf-file medical_summarization-q4_0.gguf -c 2048
```
Note: You can also use this checkpoint directly through the [usage steps](https://github.com/ggerganov/llama.cpp?tab=readme-ov-file#usage) listed in the Llama.cpp repo as well.
Step 1: Clone llama.cpp from GitHub.
```
git clone https://github.com/ggerganov/llama.cpp
```
Step 2: Move into the llama.cpp folder and build it with `LLAMA_CURL=1` flag along with other hardware-specific flags (for ex: LLAMA_CUDA=1 for Nvidia GPUs on Linux).
```
cd llama.cpp && LLAMA_CURL=1 make
```
Step 3: Run inference through the main binary.
```
./llama-cli --hf-repo fernandoruiz/medical_summarization-Q4_0-GGUF --hf-file medical_summarization-q4_0.gguf -p "The meaning to life and the universe is"
```
or
```
./llama-server --hf-repo fernandoruiz/medical_summarization-Q4_0-GGUF --hf-file medical_summarization-q4_0.gguf -c 2048
```
|